- 8.000.6071Trace-level determination of anions in the primary circuit of a PWR-type nuclear power plant using ion chromatography after inline sample preparation
The poster presents the ion chromatographic determination of organic degradation products such as glycolate, formate and acetate besides the standard anions fluoride, chloride, nitrate and sulfate.
- 8.000.6072Trace-level determination of cations in the secondary circuit of a PWR-type nuclear power plant using ion chromatography after inline sample preparation
The presented IC system with inline sample preparation allows the determination of traces of lithium and sodium (ppt) in the presence of ppm quantities of ammonium and ethanolamine.
- AB-036Half wave potentials of metal ions for the determination by polarography
In the following tables, the half-wave potentials or peak potentials of 90 metal ions are listed. The half-wave potentials (listed in volts) are measured at the dropping mercury electrode (DME) at 25 °C unless indicated otherwise.
- AB-066Potentiometric and thermometric determination of boric acid
Boric acid is used in many primary circuits of nuclear power plants, in nickel plating baths, and in the production of optical glasses. Furthermore, boron compounds are found in washing powders and fertilizers. This bulletin describes the potentiometric and thermometric determination of boric acid. The determination also covers further boron compounds, when acidic digestion is applied.
- AB-070Polarographic determination of nitrate in water samples, soil and plant extracts, vegetable juices, meat and sausages, fertilizers, liquid manure, etc.
The photometric determination of nitrate is limited by the fact that the respective methods (salicylic acid, brucine, 2,6-dimethyl phenol, Nesslers reagent after reduction of nitrate to ammonium) are subject to interferences. The direct potentiometric determination using an ion-selective nitrate electrode causes problems in the presence of fairly large amounts of chloride or organic compounds with carboxyl groups. The polarographic method, on the other hand, is not only more rapid, but also practically insensitive to chemical interference, thus ensuring more accurate results. The limit of quantification depends on the matrix of the sample and is approximately 1 mg/L.
- AB-096Determination of mercury at the rotating gold electrode by anodic stripping voltammetry
This Application Bulletin describes the determination of mercury by anodic stripping voltammetry (ASV) at the rotating gold electrode. With a deposition time of 90 s, the calibration curve is linear from 0.4 to 15 μg/L; the limit of quantification is 0.4 μg/L.The method has primarily been drawn up for investigating water samples. After appropriate digestion, the determination of mercury is possible even in samples with a high load of organic substances (wastewater, food and semi-luxuries, biological fluids, pharmaceuticals).
- AB-199Determination of sulfide and sulfite by polarography
Sulfide and sulfite can be determined polarographically without any problems. For sulfide, polarography is performed in an alkaline solution, for sulfite in a slightly acidic primary solution. The method is suitable for the analysis of pharmaceuticals (infusion solutions), wastewater/flue gas water, photographic solutions, etc.
- AB-429Determination of copper in water with the scTRACE Gold
Copper is one of the few metals which is available in nature also in its metallic form. This and the fact that it is rather easy to smelt led to intense use of this metal already in the so-called Copper and Bronze Age. Nowadays, it is more important than ever, because of its good electrical conductivity and its other physical properties. For plants and animals, it is an essential trace element; for bacteria, in contrast, it is highly toxic.This Application Bulletin describes the determination of copper by anodic stripping voltammetry (ASV) using the scTRACE Gold electrode. With a deposition time of 30 s, the limit of detection is about 0.5 μg/L.
- AN-C-049Trace cations in power plant feed water stabilized with 7 ppm monoethanolamine (MEA)
Determination of zinc, lithium, cobalt, sodium, ammonium, potassium, manganese, magnesium, and calcium in power plant feed water stabilized with 7 ppm monoethanolamine using cation chromatography with direct conductivity detection.
- AN-C-078Ethanolamines in the presence of alkali metal and alkaline earth cations
Determination of mono-, di-, and trimethanolamine (MMA, DMA, TMA respectively), in the presence of lithium, sodium, ammonium, potassium, magnesium, cesium, calcium, and strontium using cation chromatography with direct conductivity detection.
- AN-C-083Online monitoring of trace levels of cations in boiler feed water
Determination of trace levels of lithium, sodium, ammonium, potassium, magnesium, and calcium in boiler feed water using cation chromatography with direct conductivity detection.
- AN-C-090Traces of zinc and iron(II) in the presence of the standard cations
Determination of traces of zinc and iron(II) in the presence of lithium, sodium, ammonium, potassium, calcium, and magnesium in boiler water using cation chromatography with direct conductivity detection.
- AN-C-096Trace analysis of monovalent cations in an ethanolamine matrix (secondary cycle of a pressurized water reactor) using Metrohm Inline Sample Preparation with an 800 Dosino for liquid handling
Methylamine (MMA), dimethylamine (DMA), and trimethylamine (TMA) in methylpyrrolidone using Metrohm Inline Matrix Elimination.
- AN-C-126Methylamines and ethanolamines (Metrosep C 4 - 150/4.0)
Determination of monomethylamine (MMA), dimethylamine (DMA), trimethylamine (TMA), monoethanolamine (MEA), diethanolamine (DEA), and triethanolamine (TEA) using cation chromatography with direct conductivity detection.
- AN-C-132Traces of lithium, sodium, and ammonium in the presence of ethanolamine (Metrosep C 4 - 250/4.0)
Determination of lithium, sodium, ammonium, and monoethanolamine (MEA) using cation chromatography with direct conductivity detection and Metrohm Inline Preconcentration and Inline Calibration.
- AN-C-133Tap water analysis for cations using Metrohm intelligent Partial Loop Technique (MiPT)
Partial loop injection is a well known way of sample introduction to HPLC. In ion chromatography, it is not yet used to a large extent. Liquid handling with Metrohm's Dosino technology now enables to use partial loop injection on a highly reproducible and accurate level. It includes multi-level calibration out of one standard solution. This Application Note shows its use for parallel anion and cation determination in tap water applying one single Sample Processor. The anion results are shown in Application Note S–287.
- AN-C-137Copper, nickel, zinc, and common cations in the water-steam cycle of a boiling water reactor (BWR)
Water chemistry of the water-steam cycle is crucial for maintaining plant reliability and for ensuring optimal plant operational conditions. Impurities such as corrosion products in ionic, colloidal, or oxide forms are ubiquitous in feedwater, condensate, and reactor coolant. This application shows the determination of sub-ppb levels of Cu, Ni, Zn and standard cations (e.g., Na+, NH4+, Mg2+, Ca2+) in the water-steam cycle of a BWR.
- AN-C-138Zinc, nickel, calcium, and magnesium in borated water of a pressurized water reactor (PWR)
In pressurized water reactors (PWRs), light water is used as primary coolant. Boron (as boric acid) readily absorbs neutrons and is added to the coolant to control reactivity. Lithium hydroxide assures a pH value greater than 7 to prevent corrosion. This application allows to measure sub-ppb levels of zinc, nickel, calcium, and magnesium besides high boric acid and lithium hydroxide concentrations.
- AN-C-139Cations and amines in the water-steam cycle
Water in steel-based cooling systems requires a pH value slightly above 7 to prevent corrosion. Often ammonium or organic amines are applied for pH adjustement. This application shows the separation of typical amines besides inorganic cations. Sample preconcentration applies combined Inline Preconcentration and Matrix Elimination (MiPCT-ME).
- AN-C-140Lithium in borated water of a pressurized water reactor (PWR)
In pressurized water reactors (PWRs), light water is used as coolant in the primary side. Boron (as boric acid) is added to the coolant to absorb neutrons, thus controlling reactivity. Lithium hydroxide assures the alkaline pH value to prevent corrosion. This application allows to measure lithium content besides high boric acid concentrations. AN-C-138 shows the respective trace metal determination on the same system setup.
- AN-C-144Variable Inline Preconcentration including matrix elimination for trace cation determination (MiPCT-ME)
Metrohm Inline Preconcentration Technique with matrix elimination (MiPCT-ME) is a powerful method that combines preconcentration, matrix elimination, and multilevel calibration. In this Application Note, the methodology is applied to the determination of traces of sodium in addition to 2 mg/L ammonia. The Metrosep C 6 - 250/4.0 column is used for selectivity reasons.
- AN-CIC-033Monitoring PFASs in water sources
AOF (adsorbable organic fluorine) is used to screen for per- and polyfluorinated alkyl substances in aqueous matrices via pyrohydrolytic combustion and ion chromatography.
- AN-CIC-034Fast analysis of AOX in waters by CIC
Combustion ion chromatography (CIC) measures AOX (adsorbable organically bound halogens, i.e., AOCl, AOBr, AOI) and AOF as well as CIC AOX(Cl) according to DIN 38409-59 and ISO 18127.
- AN-COR-019Determining the corrosion rate with INTELLO
Tafel analysis is an important electrochemical technique used to understand reaction kinetics. By studying the Tafel slope, it reveals the rate-determining steps in electrode reactions, aiding fields like corrosion and fuel cell research. This method helps industries optimize processes and improve device performance by tailoring materials and conditions for greater efficiency.
- AN-CS-010Traces of lithium and sodium besides monoethanolamine in water-steam circuits of thermal power plants
Boiler feed water is a working medium in thermal power plant. To keep corrosion low, the pH value should be in the slightly alkali range, which is why amines are added to the feed water. This addition must be monitored regularly. Also important is the monitoring of the sodium concentration, because an increase of this indicates that cooling water is seeping into the condenser. Ion chromatography with conductivity detection following sequential suppression is the optimum system for monitoring, particularly in combination with intelligent Sample Preconcentration and Matrix Elimination.
- AN-H-012Determination of ferrous ion content of heat exchanger wash solutions
Determination of ferrous ions in heat exchanger and vessel acid wash solutions, for measuring the effectiveness of acid inhibitors used in the solutions. Depending on the condition of the sample, the lower practical limit for the determination will vary from approximately 20-100mg/Kg Fe2+. Samples with high silicic acid contents require relatively large amounts of dilution water to render them mobile, and this limits the aliquot size and hence the amount of Fe2+ which can be analyzed.
- AN-N-005Traces of silica (SiO2) in water (e.g., boiler water) after preconcentration
Determination of silica (as silicate) in pure water with preconcentration using anion chromatography with direct conductivity detection (without any post-column reaction).
- AN-N-058LOD and LOQ for silicate and borate determination according to EPA (MDL procedure)
Determination of silicate and borate and their limits of determination (LOD) and quantification (LOQ) according to the EPA procedure for method detection limit (MDL) using anion chromatography with direct conductivity detection and Metrohm Inline Calibration.
- AN-N-065Borate in borate effluent
Determination of borate in a borate effluent using anion chromatography with direct conductivity detection.
- AN-O-018Boric acid and acetic acid in process water
Determination of boric acid and acetic acid in process water using ion-exclusion chromatography with conductivity detection after chemical suppression.
- AN-P-058Amines in gas scrubber solutions using IC and pulsed amperometric detection
Scrubber solutions for scrubbing flue gas often contain amines for absorbing acid gases, e.g., sulfur dioxide (SO2). 1-(2-hydroxyethyl)piperazine and 1,4- Bis(2-hydroxyethyl)piperazine from gas scrubber solutions are separated in the Metrosep Carb 2 - 150/4.0 column and then determined using pulsed amperometric detection.
- AN-PAN-1003Online analysis of amines concentration in carbon capture plants
Carbon capture systems strip carbon dioxide from flue gases. Online analysis of amines and carbon dioxide enhances amine usage efficiency and reduces operational costs.
- AN-PAN-1013Online analysis of boric acid in the cooling water of pressurized water reactors
Boric acid requires precise monitoring in the primary circuit to control nuclear reactor reactivity. The 2060 TI Process Analyzer monitors boric acid online in near-real time.
- AN-PAN-1015Calcium and sulfate in flue-gas desulfuration
This Process Application Note deals with the online monitoring of calcium and sulfate in flue gas scrubbing solutions using titration. Other contaminants that can be measured are sulfite, chloride, and chlorine. Low concentrations of heavy metals such as cadmium, zinc, copper, and lead can be measured in the ppb/ppm range with the ADI 2045VA Process Analyzer using voltammetry.
- AN-PAN-1016Online analysis of silica in boiler feed water of power plants
Excessive silica concentrations in the boiler feed water can lead to deposits on turbine blades and must therefore be avoided. Silica analysis is carried out via differential photometry using a leading-edge technology thermostatic cuvette module for non-sample contact at the detector. Typical concentration ranges for silica are 0–50 ppb and 0–1 ppm or higher.
- AN-PAN-1032Monitoring corrosion in power plants with online process analysis
Corrosion in the water-steam circuit of power plants leads to shorter lifetimes of most metal components and potentially dangerous situations. Flow Accelerated Corrosion (FAC) is a specific case, leading to thinned pipes and elevated Fe concentrations in the circuit. Additionally, metal transport issues such as with Cu from copper heat exchangers can lead to deposition on the high pressure turbine blades, decreasing their efficiency. Current methods can monitor but not prevent these issues, and analysis times are extremely long (up to three weeks). In combination with the power plant’s Distributed Control System (DCS), online monitoring of Fe and Cu with the 2060 Process Analyzer from Metrohm Process Analytics ensures that corrosion can be controlled before it affects the power plant efficiency, ultimately decreasing downtime and lowering maintenance costs. Results are offered within 20 minutes, allowing fast adjustments to the water-steam circuit to protect company assets.
- AN-PAN-1038Power generation: Analysis of the m value (Alkalinity) in cooling water
One way to maximize heat transfer efficiency and reduce costs in a power plant is by controlling the water chemistry in the cooling circuit. This cooling water is kept alkaline to maintain the protective oxide layer on the metal piping throughout the water circuit. However alkalinity above the recommended range increases the probability of scale formation (deposition), so it is buffered with carbonate (CO32-) and bicarbonate ions (HCO3-). Titration of the cooling water to pH 4.5 gives the so-called "M-Alkalinity" (methyl orange alkalinity), a measure of total alkalinity. Below this pH, there is no more alkalinity present, only free acid (H+), carbonic acid (H2CO3), and CO2.
- AN-PAN-1040Ammonia in cooling water of thermal power plants
Thermal power plants require enormous amounts of water, using high purity steam at high pressure to rotate turbines. A separate cooling water circuit is implemented, which helps to form a vacuum when the steam condenses after the turbines. Maintaining this vacuum with optimal condensation parameters is critical for the power plant efficiency. The copper condensers are susceptible to corrosion by ammonia, leading to an upper limit of 2 mg/L NH3 set by EPRI in cooling water. Small cracks in the condenser combined with the large pressure differential between the steam circuit and the cooling water circuit will contaminate the high purity water in the boiler, causing major problems and necessitating a shutdown for plant maintenance. Monitoring NH3 online in cooling water with a process analyzer can signal early problems in a plant before significant intermediation is necessary.
- AN-PAN-1042Online trace analysis of anions in the primary circuit of nuclear power plants
Measures to monitor or prevent corrosion are crucial in nuclear power plants, where significant risks to health and safety can occur if corrosion is left unchecked. Anions corrode metals under high temperature and pressure, therefore their concentrations must be monitored at all times. The analytical challenge in the primary circuit is detection of anions in the μg/L range alongside gram quantities of boric acid and lithium hydroxide. Precise, reliable trace analysis requires the method to be automated as much as possible. The 2060 IC Process Analyzer from Metrohm Process Analytics can measure several anions from a single injection, with combined Inline Preconcentration and Inline Matrix Elimination to measure low anion concentrations precisely and reliably time after time.
- AN-PAN-1043Online trace analysis of cations in the primary circuit of nuclear power plants
In power plants, corrosion is the primary factor leading to costly and critical downtimes. In a nuclear power plant, a separate water circuit known as the Pressurized Water Reactor (PWR) ensures radioactive material stays contained while still transferring heat and energy to the other circuits. Boric acid and lithium hydroxide are added specially to the PWR circuit in amounts which can complicate other analytical measurements. Lithium prevents corrosion and must be monitored, along with other cations such as zinc, nickel, and ammonium. In order to measure these cations online at sub-µg/L range in a single analysis, the 2060 IC Process Analyzer is offered with combined Inline Preconcentration and Inline Matrix Elimination. Several cations can be analyzed in a single injection, with automated sample preparation making precise and reliable measurements easier.
- AN-PAN-1044Online trace analysis of amines in the alkaline water-steam circuit of power plants
Neutralizing amines are added to adjust pH levels within the water-steam circuit of power plants to avoid corrosion-inducing conditions. This preventive maintenance can reduce costly and critical downtimes due to corrosion, however frequent monitoring of the amine chemistry is necessary to ensure conditions stay optimal. The 2060 IC Process Analyzer featuring the Metrohm intelligent Partial Loop Technique (MiPT) option is ideal for this application, with the ability to measure trace amounts of the analytes precisely and reliably through an automated method. The benefit of using IC is that multiple analytes can be monitored simultaneously, and here the ability to measure the presence of sodium next to the high ammonium or amine concentrations could indicate that cooling water is seeping into the circuit, indicating a problem upstream.
- AN-PAN-1045Online monitoring of copper corrosion inhibitors in cooling water
Copper is used widely in industrial cooling water systems for its heat transfer properties, although it is susceptible to corrosion. Corrosion can cause a loss of efficiency and eventually a failure of equipment, leading to costly maintenance, replacement, and downtime. Corrosion inhibitors (triazoles) can be added to the water chemistry, which form sparingly soluble protective layers on the surface of the metal. Triazole concentrations must be maintained to protect the copper, which necessitates regular concentration determinations in cooling water. The 2060 IC Process Analyzer with UV/VIS detection is well-suited for this application, able to precisely and reliably measure multiple ionic and UV-active compounds simultaneously in cooling water.
- AN-PAN-1056Online monitoring of sodium in industrial power plants
In power plants, corrosion is the greatest enemy. If corrosive impurities are present in the circuit streams (e.g., chlorides and hydroxides), deposition of an insulating layer of scale on the heat transfer surfaces occurs, resulting in costly and critical downtimes. To ensure high throughput of power plants, online analysis of critical parameters such as sodium is highly advantageous for safety, protection, and process optimization. With the 2035 Process Analyzer from Metrohm Process Analytics, operators gain the information they need to accurately identify trends, reduce downtimes, and address operational issues before costly problems arise.
- AN-Q-003Online monitoring of trace levels of anions in boiler feed water
Determination of trace levels of chloride, nitrate, phosphate, and sulfate in boiler feed water using anion chromatography with conductivity detection after chemical suppression.
- AN-Q-004Online analysis of trace cations in power plant water matrices
A setup that allows online sampling is crucial for immediate and contamination-free analysis of power plant water samples. This application recommends a setup that facilitates simultaneous anion/cation determinations. Automated inline sample preparation combines variable preconcentration (MiPCT) and calibration with a single multi-ion standard. AN-Q-005 shows the respective anion results.
- AN-Q-005Online analysis of trace anions in power plant water matrices
A setup that allows online sampling is crucial for immediate and contamination-free analysis of power plant water samples. This application recommends a setup that facilitates simultaneous anion/cation determinations. Automated inline sample preparation combines variable preconcentration (MiPCT) and calibration with a single multi-ion standard. AN-Q-004 displays the respective cation results.
- AN-Q-006Online analysis of trace anions in borated water of a pressurized water reactor (PWR)
Water of the primary cycle of pressurized water reactors (PWR) contains boron for neutron absorption. The high borate content interferes with the direct analysis of trace anions. Inline Neutralization combined with variable preconcentration and Inline Matrix Elimination (MiPCT-ME) allows to remove boron as boric acid before injection.
- AN-Q-007Online analysis of chloride and sulfate in supercritical water-steam cycles
A combination of the 850 Professional IC and the 872 Extension Module Liquid Handling opens the field of Metrohm’s online monitoring by IC. In this application, Inline Preconcentration is coupled to Matrix Elimination (MiPCT-ME). By removing excess matrix components, corrosive anions can be sensitively determined. Additionally, this technique allows automated calibration using a single multi-ion standard solution. Online trace analysis for chloride and sulfate is possible for several different sample lines.
- AN-S-005Six anions in cooling water
Determination of fluoride, chloride, nitrite, nitrate, phosphate, and sulfate in cooling water using anion chromatography with conductivity detection after chemical suppression.
- AN-S-157Trace anions in boiler feed water containing 10 mg/L ammonia
Determination of traces of fluoride, chloride, bromide, nitrate, and sulfate in a boiler feed water containing 10 mg/L ammonia using anion chromatography with conductivity detection after chemical suppression and inline sample preparation by cation exchange.
- AN-S-206Online monitoring of trace anions in boiler feed water
Determination of trace levels of chloride, nitrate, phosphate, and sulfate in a boiler feed water using anion chromatography with conductivity detection after chemical suppression.
- AN-S-218Common anions in a closed cooling water system
Determination of fluoride, chloride, nitrite, nitrate, and sulfate in a closed cooling water system using anion chromatography with conductivity detection after chemical suppression.
- AN-S-222Anions in borate effluent
Determination of fluoride, chloride, nitrate, phosphate, and sulfate in a borate effluent using anion chromatography with conductivity detection after chemical suppression.
- AN-S-242Trace analysis of anions in the primary cycle of a nuclear power plant (PWR) using Metrohm Inline Sample Preparation
Determination of fluoride, glycolate, acetate, formate, chloride, nitrite, nitrate, and sulfate in the primary cycle water of a pressurized water reactor (PWR) using anion chromatography with conductivity detection after chemical suppression calibrated with Metrohm Inline Calibration.
- AN-S-295Anions in boiler water including sulfur speciation (sulfite and sulfate)
Boiler water analysis is an important task in power plant applications. Under the given conditions, the "Metrosep A Supp 10 - 100/4.0" column separates sulfite and sulfate without any organic modifier in the eluent. Even without any stabilizer, sulfite can be determined with a high reproducibility.
- AN-S-306Trace anions including chromate in water-steam cycle of a boiling water reactor (BWR)
Water of the water-steam cycle of boiling water reactors (BWR) needs to be free of corrosive anions. Analyzing these trace anions allows the parallel determination of chromate, which is a potential corrosion product. Automated sample preparation includes variable Inline Preconcentration (MiPCT) and automatic calibration with a single multi-ion calibration standard.
- AN-U-013HEDPA, PBTC, and NTP
Determination of HEDPA, PBTC, and NTP using anion chromatography with UV/VIS detection after post column reaction (PCR).
- AN-U-059Transition metals in feedwater using post-column reaction and subsequent UV/VIS detection
Feedwater for steam generation in boiling water reactors (BWR) needs to be analyzed for corrosion products. Presence of transition metals, mainly nickel and iron, indicates corrosion problems. Traces of these ions are determined using Inline Preconcentration (MiPCT). After separation, post-column reaction with 4-(2-pyridylazo)resorcinol (PAR) allows VIS detection at 510 nm.
- AN-U-060Corrosion inhibitors in cooling water
In industrial cooling water systems, copper and its alloys are widely used because of their superior heat transfer properties. These materials are, however, susceptible to corrosion. Azoles are commonly used to protect copper and its alloys from corrosion. These corrosion inhibitors are quantified by ion chromatography with UV/VIS detection.
- AN-U-061Speciation of iron(II) and iron(III) applying post-column reaction and subsequent UV/VIS detection
Speciation analysis is an important tool in analytical chemistry giving information about the quantitative distribution of different oxidation states of one and the same metal ion. The speciation of iron(II) and iron(III) (Fe 2+/Fe 3+) is achieved by ion chromatographic separation of their anionic dipicolinic acid complexes. Afterwards, post-column reaction with 4-(2-)pyridylazo-resorcinol (PAR) allows VIS detection at 510 nm.
- AN-U-078Silicate in cooling agents
Scale formation is a critical issue in cooling systems leading to system damage, which generates enormous operational losses. One important component of scale forming is silica. Amorphous silica and metal silicates especially tend to build up scale. Therefore, it is crucial to know the silica concentration in cooling agents. By ion chromatography with UV/VIS detection and PCR, it is possible to determine both free and total silicate content. Sample dilution in ultrapure water and direct injection delivers the free silicate concentration. The total silicate content is determined after hydrolysis of amorphous silica by sample dilution in alkaline eluent, and injection after a reaction time (e.g. 4 hours) prior to the analysis.
- AN-V-179Iron in boiler feed water
The iron concentration in boiler feed water has to be monitored to ensure reliable and safe operation of the water-steam circuit. Various guidelines set limits for the maximum iron content.The concentration of total iron in boiler feed water can be determined with high sensitivity using adsorptive stripping voltammetry (AdSV) using 2,3- dihydroxynaphthalene (DHN) as complexing agent. Voltammetry is a viable, less sophisticated alternative to atomic absorption spectroscopy (AAS) or inductive couple plasma (ICP) for the determination of iron with only a moderate investment in hardware required and low running costs.
- WP-056Determining dissolved oxygen in water – Titration or direct measurement?
«Dissolved oxygen» describes the amount of oxygen molecules (O2) which are dissolved in a liquid phase under certain conditions. In this white paper, two different methods for the analysis of dissolved oxygen, titration and direct measurement, are compared and contrasted to help analysts determine which method is more suitable for their specific applications. Here, we primarily focus on the determination of dissolved O2 in water. However, the same principle applies for other liquid phases such as non-alcoholic or alcoholic beverages.
- WP-062Overcoming difficulties in ion measurement: Tips for standard addition and direct measurement
Ion measurement can be conducted in several different ways, e.g., ion chromatography (IC), inductively coupled plasma optical emission spectrometry (ICP-OES), or atom absorption spectroscopy (AAS). Each of these are well-established, widely used methods in analytical laboratories. However, the initial costs are relatively high. In contrast, ion measurement by the use of an ion-selective electrode (ISE) is a promising alternative to these costly techniques. This White Paper explains the challenges which may be encountered when applying standard addition or direct measurement, and how to overcome them in order for analysts to gain more confidence with this type of analysis.
- WP-066An introduction to ion chromatography mass spectrometry (IC-MS)
Ion chromatography mass spectrometry (IC-MS) is a powerful tool that can handle many challenging analytical tasks which cannot be performed adequately by IC alone. IC-MS is a robust, sensitive, and selective technique used for the determination of polar contaminants like inorganic anions, organic acids, haloacetic acids, oxyhalides, or alkali and alkaline earth metals. After separation of the sample components via IC, mass selective detection guarantees peak identity with low detection limits. The inclusion of automated Metrohm Inline Sample Preparation (MISP) allows not only water samples, but also chemicals, organic solvents, or post-explosion residues to be readily analyzed without need for extensive manual laboratory work. This White Paper explains the benefits of IC-MS over IC in certain cases, the hyphenation of IC and different MS systems, as well as related norms and standards.
- WP-087Green alternative methods for voltammetric analysis in different water matrices
This White Paper presents four different «green» sensors: the scTRACE Gold, screen-printed electrodes, the glassy carbon electrode, and the Bi drop electrode from Metrohm that can be used to determine low concentrations of heavy metals in different sample matrices, such as boiler feed water, drinking water, and sea water.
Did you know?
Metrohm Process Analytics is the brand representing the Metrohm solution portfolio for the global process industry.