Ứng dụng
- AB-322Fully automated potentiometric determination of the hydroxyl number (HN) according to ASTM E1899 and DIN EN ISO 4629-2
The presented titration system can be used for the fully automated determination of the hydroxyl number (HN) according to ASTM E1899 and EN ISO 4629-2. The method allows, the determination of polyols and oxooils without boiling under reflux or other sample preparation and is therefore a big benefit for laboratories that have to cope with a high sample throughput. The standards EN 15168 and DIN 53240-3 relay on the same analysis method as in ASTM E1899.
- AN-T-177Hydroxyl number in binders for paints and varnishes – Pyridine-free, fully automated determination according to EN 4629-2
The hydroxyl number is an important sum parameter for quantifying the presence of hydroxyl groups in a chemical substance. As a key quality parameter, it is regularly determined in various polymers like resins, paints, polyesterols, fats, and solvents. Unlike other standards, EN 4629-2 works pyridine-free and without refluxing at elevated temperatures for a longer time. The determination is based on the catalytic acetylation of the hydroxyl group. It is performed at room temperature, requires only a small sample volumen, and can be fully automated. This Application Note describes the potentiometric determination of the hydroxyl number in 1-octanol and polyethylene glycol according to EN 4629-2. Using the OMNIS DIS-Cover technique, all sample preparation steps can be fully automated. Furthermore, the use of an OMNIS Sample Robot allows parallel analysis of multiple samples. The average time per analysis for one sample is thus reduced from approximately 49 min to 25 min., considerably increasing productivity in the laboratory.