Du er blevet omdirigeret til din lokale version af den anmodede side

Ion chromatography: Maintenance intervals

Find more information about maintenance intervals depending on the interval periods (2 months, quarterly, semi-annually, or annually), sample used, or topic (IC system, ultrafiltration, dialysis, PCR, columns, miscellaneous).

Interval – 2 months

DescriptionOrder number
Peristaltic pump tubing (orange/yellow) with three stoppers for MSM regeneration solution6.1826.420
Peristaltic pump tubing (white/white) with three stoppers to pump the sample from the Sample Processor to the injection loop6.1826.360
Peristaltic pump tubing (yellow/yellow) with three stoppers used on the sample side of the Ultrafiltration cell6.1826.390
Peristaltic pump tubing (orange/yellow) with three stoppers used on the filtrate side of the Utlrafiltration cell6.1826.320
Peristaltic pump tubing (black/black) with three stoppers used to pump the sample solution for dialysis6.1826.340
Peristaltic pump tubing (orange/yellow) with three stoppers used to pump the acceptor solution for dialysis6.1826.320
Peristaltic pump tubing (orange/green) with three stoppers used to pump the PCR reagent for the bromate determination with the triiodide method6.1826.310

Interval – Quarterly

DescriptionOrder number
Aspiration filter for the eluent
(set of 5)
6.2821.090
Spare filter for inline filter (6.2821.120) between purge valve and pulsation absorber
(set of 10)
6.2821.130
Spare filter for inline filter (6.2744.180) of the MSM6.2821.130
Metrosep C PCC 1/4.06.1010.300
Metrosep C PCC 1 HC/4.06.1010.310
Metrosep C PCC 1 VHC/4.06.1010.320
Metrosep I Trap 1 - 100/4.06.1014.200
Metrosep A PCC 2/4.06.1006.330
Metrosep A PCC 2 HC/4.06.1006.340
Metrosep A PCC 2 VHC/4.06.1006.350

Interval – Semi-annually

DescriptionOrder number
940/930/925 IC Pump
Flow check
 -  
Metrosep C Trap 1 - 100/4.06.1015.000
Metrosep A Trap 1 - 100/4.06.1014.000

Interval – Annually

DescriptionOrder number
940/930/925 IC Pump
Inlet valve PEEK
6.2824.170
940/930/925 IC Pump
Outlet valve PEEK
6.2824.160
940/930/925 IC Pump
Piston seal (orange)
6.2741.020
940/930/925 IC Pump
Zirconium oxide piston
6.2824.070

Complete glass cylinder unit, 50 mL, for Inline Eluent Preparation

Complete glass cylinder unit, 2 mL, for Dosino Regeneration

6.1574.250

6.1574.120

CO2 Absorber6.2837.100

Interval – Depending on sample

DescriptionRecommended maintenance intervalOrder number
Metrosep A Supp 4/5 Guard/4.0 (example guard)*The use of 3-4 guards per analytic column is recommended.6.1006.500
Metrosep C 4 Guard /4.0 (example guard)*The use of 3-4 guards per analytic column is recommended.6.1050.500
Metrosep A Supp 5 - 150/4.0 (example column)*2500 injections, highly sample depending6.1006.520
Metrosep C 4 - 150/4.0 (example column)*2500 injections, highly sample depending6.1050.420
Ultrafiltration membrane
(set of 50)
>100 injections, depending on sample matrix6.2714.020
Dialysis membrane cellulose (standard)
(set of 50)
40-400 injections, depending on sample matrix6.2714.010
Dialysis membrane polyamide (alternative)
(set of 50)
40-400 injections, depending on sample matrix6.2714.030
Vials PP, 11 mL (2000 pieces)For single use6.2743.050
Vials PP, 11 mL (200 pieces)For single use6.2743.057
Vials PP, 2.5 mL (2000 pieces)For single use6.2743.040
Vials PP, 2.5 mL (200 pieces)For single use6.2743.047
Stopper with perforation (2000 pieces)For single use6.2743.070
Stopper with perforation (200 pieces)For single use6.2743.077

*Caution: Number of injections can vary! It is highly depending on the sample and on the care of use!

Topic – IC system

DescriptionRecommended maintenance intervalOrder number
Peristaltic pump tubing (orange/yellow) with three stoppers for MSM regeneration solution2 months6.1826.420
Peristaltic pump tubing (white/white) with three stoppers to pump the sample from the Sample Processor to the injection loop2 months6.1826.360
Aspiration filter for the eluent
(set of 5)
quarterly6.2821.090
Spare filter for inline filter (6.2821.120) between purge valve and pulsation absorber
(set of 10)
quarterly6.2821.130
Spare filter for inline filter (6.2744.180) of the MSMquarterly6.2821.130
940/930/925 IC Pump
Flow check
semi-annually-
940/930/925 IC Pump
Inlet valve PEEK
annually6.2824.170
940/930/925 IC Pump
Outlet valve PEEK
annually6.2824.160
940/930/925 IC Pump
Piston seal (orange)
annually6.2741.020
940/930/925 IC Pump
Zirconium oxide piston
annually6.2824.070
CO2 Absorberannually6.2837.100

Topic – Ultrafiltration

DescriptionRecommended maintenance intervalOrder number
Peristaltic pump tubing (yellow/yellow) with three stoppers used on the sample side of the Ultrafiltration cell2 months6.1826.390
Peristaltic pump tubing (orange/yellow) with three stoppers used on the filtrate side of the Utlrafiltration cell2 months6.1826.320
Ultrafiltration membrane
(set of 50)
>100 injections, depending on sample matrix6.2714.020

Topic – Dialysis

DescriptionRecommended maintenance intervalOrder number
Peristaltic pump tubing (black/black) with three stoppers used to pump the sample solution for dialysis2 months6.1826.340
Peristaltic pump tubing (orange/yellow) with three stoppers used to pump the acceptor solution for dialysis2 months6.1826.320
Dialysis membrane cellulose (standard)
(set of 50)
40-400 injections, depending on sample matrix6.2714.010
Dialysis membrane polyamide (alternative)
(set of 50)
40-400 injections, depending on sample matrix6.2714.030

Topic – PCR

DescriptionRecommended maintenance intervalOrder number
Peristaltic pump tubing (orange/green) with three stoppers used to pump the PCR reagent for the bromate determination with the triiodide method2 months6.1826.310

Topic – Columns

DescriptionRecommended maintenance intervalOrder number
Metrosep C PCC 1/4.0quarterly6.1010.300
Metrosep C PCC 1 HC/4.0quarterly6.1010.310
Metrosep C PCC 1 VHC/4.0quarterly6.1010.320
Metrosep I Trap 1 - 100/4.0quarterly6.1014.200
Metrosep A PCC 2/4.0quarterly6.1006.330
Metrosep A PCC 2 HC/4.0quarterly6.1006.340
Metrosep A PCC 2 VHC/4.0quarterly6.1006.350
Metrosep C Trap 1 - 100/4.0semi-annually6.1015.000
Metrosep A Trap 1 -100/4.0semi-annually6.1014.000
Metrosep A Supp 4/5 Guard/4.0 (example guard)*The use of 3-4 guards per analytic column is recommended.6.1006.500
Metrosep C 4 Guard/4.0 (example guard)*The use of 3-4 guards per analytic column is recommended.6.1050.500
Metrosep A Supp 5 -150/4.0 (example column)*2500 injections, highly sample depending6.1006.520
Metrosep C 4 -150/4.0 (example column)*2500 injections, highly sample depending6.1050.420

*Caution: Number of injections can vary! It is highly depending on the sample and on the care of use

Topic – Miscellaneous

DescriptionRecommended maintenance intervalOrder number

Complete glass cylinder unit, 50 mL, for Inline Eluent Preparation

Complete glass cylinder unit, 2 mL, for Dosino Regeneration

annually

6.1574.250

6.1574.120

Vials PP, 11 mL (2000 pieces)For single-use6.2743.050
Vials PP, 11 mL (200 pieces)For single-use6.2743.057
Vials PP, 2.5 mL (2000 pieces)For single-use6.2743.040
Vials PP, 2.5 mL (200 pieces)For single-use6.2743.047
Stopper with perforation (2000 pieces)For single-use6.2743.070